TM 9-1240-375-34&P
GENERAL SUPPORT
M137/M137A2/M137A3 PANORAMIC TELESCOPE
(1240-01-038-0531/1240-01-483-6103/1240-01-483-6100)
FINAL INSPECTION INSTRUCTIONS
TEST AND INSPECTION
INITIAL SETUP:
Test Equipment
Azimuth test fixture (item 7, WP 0149 00)
Dioptometer (item 13, WP 0149 00)
Extension adapter (item 3, WP 0149 00)
Fixture adapter (item 5, WP 0149 00)
Infinity aiming reference collimator (item 23, WP 0149 00)
Plate-mounted level (item 12, WP 0149 00)
Surface gage (item 21, WP 0149 00)
Telescope mount holder (item 22, WP 0149 00)
Tools and Special Tools
Shop Set, Instrument and Fire Control: Field Maintenance, Basic Less Power (SC 4931-95-A07)
Tool Kit, Electronic System Maintenance (SC 5180-95-B29)
References
TM 750-116
Special Environmental Conditions
Ambient temperature +60 F (+16 C) to +90 F (+32 C)
TEST AND INSPECTION
WARNING
TRITIUM
GAS (H3)
When maintaining radioactively illuminated fire control equipment, follow radiation
hazard procedures in warning summary (M137 only).
NOTE
Procedure is written and illustrated for the M137 panoramic telescope but also
applies to the M137A2 and M137A3 panoramic telescopes.
For testing of any M137 model with quick release pins, pins must be removed to fit
azimuth test fixture.
0100 00-1
|